Electrodes
Right arm
Left arm
Right leg
Left leg
Chest 1
Chest 2
Chest 3
Chest 4
Chest 5
Chest 6
As the following figure shows, the positions of chest electrodes on the body surface are
C1: Fourth intercostal space at the right border of the sternum
C2: Fourth intercostal space at the left border of the sternum
C3: Fifth rib between C2 and C4
C4: Fifth intercostal space on the left midclavicular line
C5: Left anterior axillary line at the horizontal level of C4
C6: Left midaxillary line at the horizontal level of C4
Table 4–1 Electrodes and Their identifiers and color codes
European
Identifier
Color code
R
L
N or RF
F
C1
C2
White/yellow
C3
White/green
C4
White/brown
C5
White/black
C6
White/violet
C1
C2
C3
Identifier
Red
Yellow
Black
Green
White/red
- 27 -
Operation Preparations
American
Color code
RA
White
LA
Black
RL
Green
LL
Red
V1
Brown/red
V2
Brown/yellow
V3
Brown/green
V4
Brown/blue
V5
Brown/orange
V6
Brown/violet
C4