SE-3&SE-300 Series Electrocardiograph User Manual
3) There should be no shield in or around the room where the wireless AP is used, or
else the wireless transmission may fail.
 Auto Transmission:
1. Press the MENU key to open the System Setup Screen.
2. Select the Transmission button, and then press the MENU key to enter the sub-window.
3. Set Auto Transmission to On and set Transmission mode to Ethernet.
4. Set the Server IP to the IP of Smart ECG Viewer.
5. Set the first three numbers of the Local IP to the first three numbers of the IP of Smart
ECG Viewer. The last number of the Local IP item can be set at random, but it can't be the
same as the last number of the IP of Smart ECG Viewer.
For example, check the IP on the System Setting interface of the Smart ECG Viewer
software.
Set the Server IP to the IP of Smart ECG Viewer. Set the first three numbers of the Local
IP to the first three numbers of the IP of Smart ECG Viewer. The last number of the Local
IP can be set at random, but it can't be the same as the last number of the local IP of Smart
ECG Viewer.
6. In the AUTO, RHYT, OFF-AUTO or OFF-RHYT mode, ECG data can be transmitted
through the net automatically after an ECG report is printed out.
 Manual Transmission:
1. Press the MENU key to open the System Setup Screen.
2. Select the Transmission button, and then press the MENU key to enter the sub-window.
3. Set Auto Transmission to Off and set Transmission mode to Ethernet.
4. For IP address setting, refer to step 4 and 5 for auto transmission.
5. Other steps for transmission operation, refer to step 4 for manual transmission in section
7.1.1 "Transmitting ECG Record through the Serial Port".
NOTE: The transmission process is long, and please be patient to wait.
- 41 -
Managing ECG Records