SE-3&SE-300 Series Electrocardiograph User Manual
Lowpass Filter
Rhythm Lead1/2/3
Lead Sequence
Lead
Sequence
Standard І, II, III
Cabrera
NOTE: To pass the distortion test, the electrocardiograph has to be configured with the
highest bandwidth in filter settings. Otherwise, ECG signal may be distorted.
The cutoff frequency can be set to 75Hz, 100Hz or 150Hz, the default
value is 100Hz
Lowpass Filter restricts the bandwidth of input signals.
All the input signals whose frequency is higher than the set cutoff
frequency will be attenuated.
NOTE: Only when EMG Filter is set to Off, can the setting of
Lowpass Filter be effective.
Choose from: І, П, III, aVR, aVL, aVF, V1, V2, V3, V4, V5, or V6, the
default value is of Rhythm Lead 1 is II, the default value is of Rhythm
Lead 2 is V1, the default value is of Rhythm Lead 3 is V5
In the auto mode:
When Record Style is set to 3× 4+1R or 6× 2+1R, the rhythm lead
selected in the Rhythm Lead1 list box will be printed in the ECG reports;
In the rhythm mode:
When Rhythm Style is set to Single Lead, 60s wave of the rhythm lead
selected in the Rhythm Lead1 list box will be printed in the ECG reports;
When Rhythm Style is set to Three Leads, 20s waves of three rhythm
leads selected respectively in the Rhythm Lead1/2/3 list box will be
printed in the ECG reports.
In the R-R analysis mode:
The R-R analysis report of the rhythm lead selected in the Rhythm Lead1
list box will be printed.
NOTE: Rhythm Lead 1/2/3 must be different from each other.
Choose from: Standard or Cabrera, the default value is On
Lead group 1
Lead group 2
aVR, aVL, aVF V1, V2, V3
aVL, І, -aVR
II, aVF, III
Lead group 3
V1, V2, V3
- 47 -
Settings
Lead group 4
V4, V5, V6
V4, V5, V6