Functional Safety
10.4 Disabling Functional Safety over device with display
5. Validate the identification data of your device again: Long tag, product name and serial
number.
6. Click "Ok".
Functional Safety is disabled.
The device restarts automatically.
Result
● The safety-related error is acknowledged.
● The device returns to the "Functional Safety disabled" device mode.
Note
The system shows unexpected data or behaves differently than described in this procedure.
● Repeat the entire procedure.
10.4
Disabling Functional Safety over device with display
Requirement
The device is in the "Functional Safety enabled" device mode.
170
SITRANS P320/P420 (mA/HART)