Operating
7.2 Operating the device with display
7.2.3.2
Navigating in the measurement view
Requirement
You have disabled the button lock.
Disabling button lock (Page 131)
Procedure
1. Use the
2. To switch to the parameter view, press the
7.2.4
Parameter view
The parameter view shows the parameters, parameter values and the wizards of the device.
①
②
③
④
Figure 7-4
For parameters with an associated unit, the parameter name and unit are displayed as
alternating values in
7.2.4.1
List of parameters on the display
The parameters are displayed with parameter ID and parameter name.
Depending on the parameter settings of your device, some parameters are not visible.
Parame‐
Parameter name on
ter ID
the display
01
PRESS UNITS
02
LOWER RANGE
03
UPPER RANGE
04
DAMPING
05
APPLICATION
74
and
buttons to navigate in the measurement view.
Name and unit of the parameter (alternating)
Parameter value
Parameter ID
"EDIT" symbol (permanently enabled)
Example of parameter view
①
. Example: Lower range value in mA.
Meaning
Pressure units (Page 101)
Set lower range value (without pressure applied) (Page 104)
Set upper range value (without pressure applied) (Page 104)
Damping value (Page 105)
Application
(Page 106)
1)
button.
SITRANS P320/P420 (mA/HART)
Operating Instructions, 06/2018, A5E44852162-AA