Apply Lower Range Value [08]/Apply Upper Range Value [09]; Apply Lower Range Value Parameter [08]; Apply Upper Range Value Parameter [09] - Siemens SITRANS Series Operating Instructions Manual

Pressure transmitter ma/hart
Hide thumbs Also See for SITRANS Series:
Table of Contents

Advertisement

Parameter assignment
9.2 Parameter assignment over device with display
Procedure
1. In the parameter view, select the "Zero point adjustment" parameter [07].
2. Navigate into the parameter view.
Navigating in the views (Page 71)
3. Press the
4. The value "0" appears on the display.
5. Confirm the value by pressing the
6. Change to the measurement view with the
Result
The device displays pressure measurement 0.0 in the set unit.
Depending on the set damping, the settling time is extended until the pressure measurement
0 is displayed.
Note
For devices for absolute pressure, the lower range value is at vacuum (0 bar a).
The zero point adjustment for devices for absolute pressure that do not measure absolute
pressure (0 bar a) leads to incorrect settings.
Note
The effective measuring range is reduced by the amount of the upstream pressure.
Example: With an upstream pressure of 10 mbar, the effective measuring range of a 1300 mbar
pressure transmitter is reduced to a point between 0 and 1290 mbar.
9.2.7

Apply lower range value [08]/Apply upper range value [09]

9.2.7.1

Apply lower range value parameter [08]

Sets the lower range value to the current reference pressure.
Setting range:
Factory setting:
9.2.7.2

Apply upper range value parameter [09]

Sets the upper range value to the current reference pressure.
Setting range:
Factory setting:
118
button.
Within the measuring limits
See nameplate (depending on measuring cell)
Within the measuring limits
See nameplate (depending on measuring cell)
button.
button.
Operating Instructions, 06/2018, A5E44852162-AA
SITRANS P320/P420 (mA/HART)

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

Sitrans p320Sitrans p420Sitrans p series

Table of Contents