Maintenance; Maintenance Interval; Gasanalyzer - Emerson Rosemount BINOS 100 F Instruction Manual

Addendum for pressurized analyzers intended to be used in hazardous areas classified zone 2
Hide thumbs Also See for Rosemount BINOS 100 F:
Table of Contents

Advertisement

ATEX Instruction Manual
HAS8xE-IM-ATEX
January 2004
After maintenance or replacement of parts concerning explosion protection an
authority on explosion protection has to verify that the analyzer still meets the
requirements for explosion protection before it is switched on again.
If parts essential for explosion protection are repaired they have to be routine
tested!
The authority has to issue a certificate for this and/or attach a test label to the
equipment before startup after maintenance or replacement of parts.
4-1

Maintenance Interval

To ensure the performance and safety of the
equipment it has to be checked on a regular
basis, at least once a year. Special care has
to be taken for the EEx p saftey device and
parts ensuring explosion protection (e.g.
gaskets).
4-2

Gasanalyzer

Refer to the associated manual for detailed
information on maintenance, replacement of
parts and how to carry out a containment
system leak test.
Emerson Process Management Manufacturing GmbH & Co. OHG
SECTION 4

Maintenance

MLT 2 / BINOS
It is the operators/owners
responsibility to extend the
maintenance interval with
respect
to
influences of gases or
environment on materials in
contact with the sample gas
or
ensuring
protection (e.g. gaskets).
100 F
negative
explosion
4 - 1

Advertisement

Table of Contents
loading

This manual is also suitable for:

Rosemount mlt 2

Table of Contents