Additional Safety Measures - Emerson Rosemount BINOS 100 F Instruction Manual

Addendum for pressurized analyzers intended to be used in hazardous areas classified zone 2
Hide thumbs Also See for Rosemount BINOS 100 F:
Table of Contents

Advertisement

MLT 2 / BINOS
1-4

Additional Safety Measures

Exhaust gas lines must end outside the
hazardous area at a safe point. Sample
gas lines may be returned to the
sampling point.
If gas exhaust lines end within the
hazardous area gas inlet and outlet
have to be equipped with flame
arrestors.
The maximum permissible gas
pressure depends on the type of
connected gas. See section 1-3-5 for
detailed information.
1 - 6
100 F
1- Additional Safety Measures
ATEX Instruction Manual
Emerson Process Management Manufacturing GmbH & Co. OHG
HAS8xE-IM-ATEX