MLT 2 / BINOS

2-7-2-2 Analyzer

The mains power line cord has to be
connected to the terminals provided at the
analyzer enclosure's inner left side, near the
EMC line filter (fig. 2-5).
Fig. 2-5
Terminal location inside analyzer
By default the analyzer has to be supplied
separatly to the pressurization unit. This
Omitting the pre-purge sequence
The pre-purging sequence can be refrained
if it is ensured that at the moment of
activating the analyzer the atmosphere
inside the enclosure and the appropriate
pipes is well below 25 % LEL.
2 - 16
100 F
2-7-2 Mains Connection
ATEX Instruction Manual
Verify that the mains voltage
at the point of installation
meets the analyzer nominal
voltage!
EMC line filter
Terminals
allows to shut on the safety device for pre-
purging the analyzer, while this is still shut
off!
It is the responsibility of the
user to omit pre-purging!
Emerson Process Management Manufacturing GmbH & Co. OHG
HAS8xE-IM-ATEX
January 2004