13. Specifications
Electrical Voltage
Current
Phase
Frequency
Water Pressure
Water Temperature
Water Flow Rate
Weight
Dimensions
Footswitch
Remote Communication
Operating Environment
Transport and Storage Conditions
Continuous (100-240 VAC)
1.0 Amperes, Maximum
Single
50/60 Hertz
20 to 40 psig (138 to 275 kPa)
< 25°C (77°F)
Minimum Setting (CCW) < 15 ml/min
Maximum Setting (CW) > 55 ml/min
4.4 lbs (2 Kg)
Height: 3.7 in (9.398 cm)
Width: 6.9 in (17.526 cm)
Depth: 10.7 in (27.178 cm)
Handpiece Cable length: 6.5 ft (2.0 M)
Auxillary Footswitch Cable length: 6 ft (1.8 M)
Water Supply Line length: 8 ft (2.4 M)
Protection Class IPX1.
Not for operating theatres.
Frequency: 2.4 GHz
Power: < 1mW
Temperature:
15 to 40 Deg Celsius
(59 to 104 Deg Fahrenheit)
Relative Humidity:
30% to 75% (non-condensing)
Temperature:
40 to 70 Deg. Celsius
(-40 to 158 Deg. Fahrenheit)
Relative Humidity:
10% to 95% (non-condensing)
Atmospheric Pressure: 500 to 1060 hPa
English - 13