Download Print this page

Emerson AMS 2140 Quick Start Manual page 34

Machinery health analyzer advanced laser alignment application
Hide thumbs Also See for AMS 2140:

Advertisement

Quick Start Guide
2. Set the connection type on the AMS 2140.
3. Enable the AMS 2140 in Data Transfer.
4. Set the connection type in Data Transfer. The connection type on the
AMS 2140 and on Data Transfer must match.
Refer to the AMS 2140 Machinery Health Analyzer User Guide for more
information on setting the connection type on the AMS 2140 and on AMS
Machinery Manager Data Transfer.
Procedure
1. From the Laser Align Application main menu, press ALT > F9 Job Manager >
F7 Connect for Transfer.
2. Optionally, press F1 Set Storage Location to select the internal memory or
an SD memory card.
This option is available only if a memory card is inserted into the AMS
2140.
3. Press F7 Load Jobs.
4. Press F8 or F9 to select a database.
5. Press F10 or F11 to select an area within the database and press F4 Select
Area.
6. Press F8 or F9 to select an equipment within the area and press F7 Select.
7. Press F8 or F9 to select a job and press F1 Select.
8. Press F7 Load Jobs.
5.3
Load tolerance table from AMS Machinery Manager
Alignment tolerance tables can only be transferred from the PC to the AMS
2140.
The alignment tolerance table contains the tolerance values for standard and
jackshaft tolerances as well as soft foot and frame distortion index (FDI)
tolerance values for horizontal alignment jobs.
Prerequisites
1. Connect the AMS 2140 to a PC with AMS Machinery Manager installed
and running.
2. Set the connection type on the AMS 2140.
3. Enable the AMS 2140 in Data Transfer.
4. Set the connection type in Data Transfer. The connection type on the
AMS 2140 and on Data Transfer must match.
34
AMS 2140 Advanced Laser Alignment Application
December 2018

Hide quick links:

Advertisement

loading