Quick Start Guide
3. Use the touchscreen or keypad to select an entry from the Machine Type
list describing the left machine and press F5 Select Left Machine.
4. Use the touchscreen or keypad to select an entry from the Machine Type
list describing the right machine and press F11 Select Right Machine.
5. Optionally, press F8 Change Machine View to change the machine view from
2D to an isometric machine view.
6. Press Enter.
3.3.3
Set the alignment method
Selecting an alignment method is available only in Advanced mode. In Basic
mode, Manual 4 Point is the only available alignment method.
Procedure
1. Activate a horizontal alignment job.
2. From the Laser Align Application main menu, press ALT > F4 Laser Align
Method.
3. Select an alignment method.
The Basic Laser Alignment Application defaults to Manual 4 Point. The
Advanced Laser Alignment Application lets you select from 5 different
methods.
This document only covers Auto Sweep and Manual 4 Point. Refer to the
Alignment chapter of the AMS 2140 Machinery Health Analyzer User
Guide for more information on the other available alignment methods.
a. For Auto Sweep, also select Standard or Averaging as the mode of
operation.
b. For Manual 4 Point, also enter a sample rate.
The sample rate sets the number of samples (1 to 25) to be averaged
to generate a single reading.
4. Press Enter.
3.3.4
Change the laser and sensor placement
Change the laser and sensor placement on the AMS 2140 to reflect your
current laser and sensor setup.
Changing the placement of the laser and sensor is available only in Advanced
mode.
Procedure
1. Activate a horizontal alignment job.
2. From the Laser Align Application main menu, press ALT > F8 Head Config.
18
AMS 2140 Advanced Laser Alignment Application
December 2018