DSK-1031 is intended for use in the electromagnetic environment specifi ed below. The customer or the user of DSK-1031 should assure that it is used in such an environ-
ment.
Immunity test
Electrostatic discharge (ESD)
IEC 61000-4-2
Electrical fast transient/burst
IEC 61000-4-4
Surge IEC 61000-4-5
Voltage dips, short interruptions
and voltage variations on power
supply input lines IEC 61000-4-11
Power frequency (50/ 60 Hz)
magnetic fi eld IEC 61000-4-8
NOTE: U
is the a.c. mains voltage prior to application of the test level.
T
Table 202. Guidance and manufacturer's declaration - electromagnetic immunity
IEC 60601 test level
±6 kV contact
±8 kV air
±2 kV for power supply lines
±1 kV for input/output lines
±1 kV diff erential mode
±2 kV common mode
<5% U
(>95% dip in U
) for 0,5
T
T
cycle
40% U
(60% dip in U
) for 5
T
T
cycles
70% U
(30% dip in U
) for 25
T
T
cycles
<5% U
(>95% dip in U
) for 5 sec
T
T
3 A/m
Compliance level
Electromagnetic environment – guidance
±6 kV contact
Floors should be wood, concrete or ceramic tile. If fl oors are
±8 kV air
covered with synthetic material, the relative humidity should
be at least 30 %.
±2 kV for power supply lines
Mains power quality should be that of a typical commercial
±1 kV for input/output lines
or hospital environment.
±1 kV diff erential mode
Mains power quality should be that of a typical commercial
±2 kV common mode
or hospital environment.
<5% U
(>95% dip in U
) for 0,5
Mains power quality should be that of a typical commercial
T
T
cycle
or hospital environment. If the user of DSK-1031 requires
40% U
(60% dip in U
) for 5
continued operation during power mains interruptions,
T
T
cycles
it is recommended that DSK-1031 is be powered from an
70% U
(30% dip in U
) for 25
uninterruptible power supply or a battery.
T
T
cycles
<5%U
(>95% dip in U
) for 5
T
T
sec
3 A/m
Power frequency magnetic fi elds should be at levels
characteristic of a typical location in a typical commercial or
hospital environment.