follow these steps:
(1) Check whether the probes are correctly connected to the signal line connection.
(2) Check whether the signal cable is properly connected to the BNC.
(3) Check whether the probe is properly connected with the analytes.
(4) Check whether the analyte signal is generated.
(5) Re-acquire the signal again.
3. Measured voltage amplitude value is 10 times greater than the actual value,
or 10 times smaller:
Check whether the channel attenuation factor of the probe matches the actual
attenuation ratio.
4. There waveform display, but not stable:
Check the trigger source trigger menu settings are consistent with the actual
signal input channels.
5. Press "RUN" button without any display:
Trigger checks whether the trigger menu in the "normal" or "Single" and the
trigger level has been exceeded if the signal range. If it is, the trigger level is
centered, or set the trigger mode to "AUTO" file.
6.Stepped waveform display:
This phenomenon is normal. When the base level is too low may stall,
increasing the level of the base when the horizontal resolution can be
increased to improve the display.

Chapter 5 service and support

5.1 Warranty Description

We guarantee the production and sale of its products, the date of shipment from
authorized dealers within a year, does not appear in material and workmanship defects.
As specified in the detailed product warranty proved defective, we will provide repair or
replacement service.
In addition to this summary, or use the warranty provision of the warranty, we do not make
any other warranties, express or implied. The Company's indirect, special or damage
arising there from shall not be liable.

Appendix A: Technical Specifications

Unless otherwise noted, all technical specifications are used for attenuation switch setting
10X probes and these series oscilloscopes.
To verify that the oscilloscope meets specifications, the oscilloscope must meet the
following conditions:
●The oscilloscope must be more than thirty minutes of continuous operation within the
37