3, the measured voltage amplitude value is 10 times greater than the actual value or small 10
times:
Check whether the channel attenuation coefficient is consistent with actual use of probes
attenuation ratio.
4, a waveform display, but can not be stabilized:
Check whether triggering menu of trigger source set is consistent with the actual signal input
by the channel.
5, press the "RUN" button without any display:
Check whether the trigger menu trigger mode is in a "normal" or "single", and the trigger level
is beyond waveform range. If so, will trigger level center, or set the trigger mode to "AUTO".
6, waveform display ladder shaped:
This phenomenon is normal. May level when the gear is too low, increase the level of time base
can improve horizontal resolution, improves the display.