Keysight N437x Series Lightwave Component Analyzer, Programmer's Guide
:PARameter:OPTical:SOURce:S2PFile
command:
:PARameter:OPTical:SOURce:S2PFile<wsp>"<path string>"
syntax:
Specifies the name of the s2p file to use for additional adaptor de-
description:
embedding on the source side. In default mode only evaluated for O/E
and O/O measurements.
"<string>" the file name and path enclosed in double quotes.
parameters:
response:
<string> file name and path of the s2p file on the LCA system.
C#:
(property) OptSrcFile
example:
:PAR:OPT:SOUR:S2PF "c:\temp\test1.s2p"
:PARameter:RECeiver:ATTenuation?
command:
:PARameter:RECeiver:ATTenuation?
syntax:
Retrieves the optical attenuation on the receiver path. In default mode
description:
only evaluated for E/O and O/O measurements
none
parameters:
<double> attenuation value, default unit is dB
response:
C#:
(property) RcvAttOpt_dB
example:
:PAR:REC:ATT? -> 3
:PARameter:RECeiver:ATTenuation
command:
:PARameter:RECeiver:ATTenuation <wsp><attenuation>
syntax:
Specifies the optical attenuation on the receiver path. In default mode only
description:
evaluated for E/O and O/O measurements
parameters:
attenuation <double> attenuation value, default unit is dB
none
response:
C#:
(property) RcvAttOpt_dB
example:
:PAR:REC:ATT 2
:PARameter:RECeiver:PATH:LENGth?
command:
:PARameter:RECeiver:PATH:LENGth?
syntax:
Retrieves the geometrical length of the receiver path in m. In
description:
default mode only evaluated for E/O and O/O measurements
parameters:
none
response:
<double> The path length value, default unit is meter.
C#:
(property) RcvLengthOpt_m
example:
:PAR:REC:PATH:LENG? -> 0.3
:PARameter:RECeiver:PATH:LENGth
command:
:PARameter:RECeiver:PATH:LENGth<wsp> <length>
syntax:
Specifies the geometrical length of the receiver path in m. In
description:
default mode only evaluated for E/O and O/O measurements
length <double> path length value, default unit is meter.
parameters:
response:
none.
C#:
(property) RcvLengthOpt_m
example:
:PAR:REC:PATH:LENG 0.45
command:
:PARameter:RECeiver:REFR:INDex?
:PARameter:RECeiver:REFR:INDex ?
syntax:
Retrieves the refractive index of the receiver path in dB. In default
description:
mode only evaluated for E/O and O/O measurements.
none
parameters:
<double> the refractive index value, unit is dB.
response:
(property) RcvRefIdx
:PAR:REC:REFR:IND? -> 0
example:
Remote Operation