Setting Up Serial Communications - GE DigitalFlow GS868 Programming Manual

Panametrics steam ultrasonic mass flowmeter, 2-channel
Hide thumbs Also See for DigitalFlow GS868:
Table of Contents

Advertisement

March 2008
Setting Up Serial
Communications
1-42
The Model GS868 flowmeter can transmit stored data and displayed
readings to a remote ANSI terminal or a personal computer by
connecting the meter's RS232 interface to the serial port of the PC. In
addition, the Model GS868 can receive and execute remote
commands via this link.The meter can also be configured with a
MODBUS option card for MODBUS (RS485) communications, a
Foundation Fieldbus option card for Foundation Fieldbus
communications, an Ethernet option card for Ethernet
communications, or with a MODBUS/TCP option card for MODBUS
communications over the Ethernet, if desired.
Note: To set Fieldbus communications parameters, see Appendix D,
Foundation Fieldbus Communications.
COMM
Use the
submenu to set the RS232, MODBUS, Ethernet and
MODBUS/TCP communications port parameters and to enter a
network identification number.T A network identification number is
required to use the GE Instrument Data Manager or PanaView
software. While following the programming instructions, refer to the
menu map in Figure A-3 on page A-3.
1. To enter the
COMM
submenu, press
screen.
2. Use the [←] and [→] and
Rate. The available rates are 300, 600, 1200, 2400, 4800, 9600 and
19,200 baud.
3. Use the [←] and [→] and
UART Bits, as shown in Table 1-11 below.
Table 1-11: UART Bits Options
Option Bar
# Data Bits
8,no
8,odd
8even
7,odd
7even
4. Enter a Network ID number between 1 and 254 and press
The default number is 1.
A network ID number is only necessary for communication with the
GE Instrument Data Manager or PanaView software. See the
software User's Manual for more information.
IMPORTANT:
If the network ID number is changed, communication
with the Instrument Data Manager or PanaView must
be re-established with the new ID number.
[F4]
Global PROGRAM
at the
[F1]-[F4]
keys to select the desired Baud
[F1]-[F4]
keys to select the number of
# Stop Bits
8
0
8
0
8
0
7
1
7
1
Programming Site Data
Parity
None
Odd
Even
Odd
Even
[ENT]
.

Advertisement

Table of Contents
loading

Table of Contents