5 Troubleshooting

Malfunction
LED display
Device does
not start
Device does
not charge
Diminished
flow rate
Insufficient
flow rate
No flow rate
24
Probable causes
LEDs do not
Battery is discharged
illuminate
Battery is discharged and impermissible
LED blinks red
input voltage
Temperature of the battery too high for
LED blinks red
charging
LED illuminates
Battery is defective
green
LED blinks red
Impermissible input voltage
LED blinks
Device is overheated and is only running in
green/yellow
curbed operation.
LED illuminates
External bacterial filter blocked
green/yellow
Integrated bacterial filter blocked
Leak in the tubing system or in the canister
LED illuminates
system
green
Float ball of the overflow protection blocks
the aspiration lid inlet
LED illuminates
green/yellow
Hydrophilic overflow protection of the dis-
posable secretion canister system blocked
External bacterial filter blocked
LED illuminates
Integrated bacterial filter blocked
green
Suction tube blocked
Remedy
Connect the power supply unit for charging.
Connect the correct power supply unit,
check the power supply.
No measure required, automatic resumption of charging
when battery has cooled down.
Please contact service!
Connect the correct power supply unit,
check the power supply.
Switch off the device and allow it to cool down.
Replace the external bacterial filter
Please contact service!
Check that secretion canister lid and tubing connections
are tight and no leakages are apparent.
Check fluid level in the secretion canister, empty the
canister
Replace the disposable liner 'OneWay'
Replace the external bacterial filter
Please contact service!
Replace suction tube