1.0 INTRODUCTION INTRODUCTION CRW Series offers heavy-duty scales used to weigh very large masses. The main features include- • Large LED display (30.5 mm) • Rechargeable battery operated • Selectable reading resolution • Charger auto-off • Battery protection • Standard Remote Control Handset...
Stable indicator- this indicator means the current data is stable and you can read the data. When in calibrate mode, make sure to proceed only when this indicator is on. Zero indicator- this indicates the scale is in zero point. @Adam Equipment Company 2004...
Safety Lever returns to its resting position which prevents the sample from slipping out of the hook. • Do not exceed the maximum capacity of the Scale as mentioned in section 1.2. • Clean and protect your scale after use. @Adam Equipment Company 2004...
Overload Warning- Please do not add any item that exceeds the maximum capacity of the scale. When the display shows , remove the item to avoid damage to the load cell. @Adam Equipment Company 2004...
After entering the value of the calibration mass, press the [ZERO] key to confirm. When display shows “ ”, add the calibrate mass on the hook after stable indicator on. Press the [ZERO] key again, wait for few seconds, span calibration is now complete. @Adam Equipment Company 2004...
Page 8
4.2 PARAMETER SETTING TABLE FUNCTION DESCRIPTION DEFAULT Sets division : high resolution : standard resolution : low resolution Sets zero tracking range Sets baud rate (RS-232 is an optional function) : 600bps, : 1200bps :2400bps, : 4800bps : 9600bps @Adam Equipment Company 2004...
Page 9
: auto sleep after scale empty for 20 minutes : auto sleep after scale empty for 30 minutes Sets parameter setting password Calibration " " $ % (calibration mass) " $ % (complete) For details see SECTION 3. @Adam Equipment Company 2004...
Remove weight from the scale if overloaded. The load cell or the electronics may be faulty. If the problem persists, contact your dealer or Adam Equipment for assistance. Battery low Charge the battery immediately. @Adam Equipment Company 2004...
Page 11
Shielded interconnect cables must be employed with this equipment to insure compliance with the pertinent RF emission limits governing this device. Changes or modifications not expressly approved by Adam Equipment could void the user' s authority to operate the equipment.
Page 12
Copyright by Adam Equipment Co. Ltd. All rights reserved. No part of this publication may be reprinted or translated in any form or by any means without the prior permission of Adam Equipment. Adam Equipment reserves the right to make changes to the technology, features, specifications and design of the equipment without notice.