GSP9700 Series Road Force Measurement System Operation
Position the calibration tool parallel to the floor. Tap the foot pedal once or press
"OK."
NOTE:
To verify that the tool is parallel to the floor, measure from
both ends of the calibration tool to the floor. The
measurement will be the same at both ends when the tool is
parallel to the floor.
®
Place the inner Dataset
"Enter Cal Step."
WHEN POSITIONING THE INNER DATASET ARM,
VERIFY THAT THE INDENT OF THE ARM IS
CENTERED
ON THE BUTTON FOR EACH POSITION.
NOTE:
If the inner Dataset
a long, high-pitched tone will sound to advise that the step
has NOT been entered. Stabilize the arm and re-enter the
step.
®
Place the inner Dataset
"Enter Cal Step."
®
Place the inner Dataset
"Enter Cal Step."
arm at upward position "1." Tap the foot pedal once or press
®
arm is unstable while entering any step,
arm at upward position "2." Tap the foot pedal once or press
arm at upward position "3." Tap the foot pedal once or press
3
2
1
7. Calibration and Maintenance • 119