Emerson Rosemount 2130 Reference Manual page 18

Level switch
Hide thumbs Also See for Rosemount 2130:
Table of Contents

Advertisement

Installation
January 2019
As a guideline, the maximum solid particle diameter in the liquid is 0.2 in. (5 mm).
Extra consideration is needed when dealing with particles bigger than 0.2 in. (5 mm).
Consult the factory for advice.
In almost all cases, the level switch is insensitive to foams (i.e. does not see the foam).
However, on rare occasions, some very dense foams may be seen as liquid. Known examples of this are
found in ice-cream and orange juice manufacturing.
Figure 2-3. Switching Points
SP
SP
SP. Switchpoint (H
HY. Switching Hysteresis
Note
When mounted vertically, a low density media has a switching point closer to the process connection.
A high density media has a switching point closer to fork tip.
10
0.5 in.
(13 mm)
0.5 in.
(13 mm)
0)
2
0.1 in.
(2.5 mm)
HY
Reference Manual
00809-0100-4130, Rev DC
Installation

Advertisement

Table of Contents
loading

Table of Contents