Alcatel Drytel 100 Instruction Manual page 17

Pumping unit
Table of Contents

Advertisement

Relubrication process:
1) Bearing opposite the pumping unit (see figure 1) :
- Unscrew the regreasing plug, (7B)
- Position the syringe in the orifice perpendicular to axis of rotation
- Remove one spacer and inject progressively the grease. The piston must touch the ring spacer (or the other
black spacer).
- Remove the syringe.
- Change the seal, (8B), and reinstall the plug.
the
2) Bearing nearer
pumping unit (see figure 1) :
- Unscrew the regreasing plug,
- Position the syringe in the orifice inclined at a 45 degrees angle to the axis of rotation.
- Remove the other black spacer and inject the grease into the bearing.
- Remove the syringe.
- Change the seal, (SA), and reinstall the plug.
Reinstall the MDP on its shocks absorber in reverse order from parag. 5.2.1.
Connect the power line, the vacuum line, and water line.'
5.2.3 - ST ART UP AFTER REGREASING
1) Run the MDP at atmospheric pressure without cooling for about
- Start the DRYTEL, the inlet opened.
If the red fault lamp comes on before the MDP has run
shut off the pump. In this case, wait for the red lamp to go off then restart the DRYTEL. Repeat this
procedure until the MDP runs continuously for
2) Now run the MDP with cooling and with the backing pump according to the following procedure, the
inlet closed :
RUNNING PROCEDURE
1 min 15 s
42
(7A).
5
minutes.
min
3
34 min 15 s
� •.
5
minutes as follows :
5
minutes it indicates that
15 min
· ----
thi
protection switch has
r---
1
I
15 mln ----1
t
s.:
--
.,.
- be
- 511!
- Sf:
- in
.!.
SE1
,._
::
""(
Exe
- 1 :
-2
* l

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents