INTERFACE (Options) (P 104)
This plug allows
. The remote starting stopping of the ORYTEL.
. The output by d
contact closures, from the state of the three lamps (red, yellow, green ) , from the
ry
"ON-OFF"
state of the converter, from the
option, see figures 7 and 8 for the electrical connections.
If the remote control is not used, strap the terminals 22 and 9 on Pl 04 socket (see figure 8).
3.4 - STARTING
For the initial start-up, or when the pump has not been used for a 2 month period, it is recommended that
the MOP be operated for about 5 minutes at atmospheric pressure. This procedure is necessa
redistribute the grease in the ball bearings.

3.5 - APPLICATIONS NOTES

3.51. The membranes pump must be vented to atmosphere whenever the Orytel is stopped.
To accomplish this an automatic vent valve is employed. This valve vents the membranes pump while
isolating the MOP and the chamber whenever Orytel is stopped.
3.52. The membranes pump must never be used to pump liquids.
3.53. Orytel must be sheltered from rain, leaking water and excessive dust.
3.54. If the foreline or exhaust is restricted in any way then care must be taken to be sure that the maximum
foreline pressure given in the specifications is not exceeded. Excessively high foreline pressure will cause
the MOP to shut off after 8 minutes. (See 1.21 b)
I
3.55. Warning
For the transport, the membranes pump is fastened on its frame. Before start the Drytel, proceed like this.
- Disconnect the line cable
- Uscrew the 4 assembling screws located on the side panels of the ORYTEL and remove the body.
- Disconnect the electrical cable from the frame (on Jo connector) (see figure 7).
- Unscrew the 2 assembling nuts from the clamping strap and remove it, with its to connection pins.
- Reconnect the electrical cable to connector Jo.
- Install and secure with screws the body on the frame.