INFINITY DELTA/DELTA XL/KAPPA - QUICK REFERENCE GUIDE
RESP Problems
RESP Coincidence
RESP Can't detect
coincidence
RESP Signal
Saturated
Infinity Delta/Delta XL/Kappa - Quick Reference Guide
Intended for quick reference only.
Refer to Instruction For Use Manual.
Possible Cause
Patient's heart rate
and respiration rate fall
within 20% of each
other.
RESP Coincidence is
enabled but there is
excessive ECG artifact.
Or ECG leads are off.
RESP signal detected
by monitor has
excessive baseline
shift.
Action
Observe the patient
carefully. Treat if nec-
essary. Check and or
change the electrode
placement if you
receive a coincidence
message until you
obtain a clear respira-
tion signal.
Calm patient. Ensure
that the patient's skin
is properly prepped
and change electrodes
if necessary. Isolate the
patient form auxiliary
equipment. Inspect and
replace defective
cables and wires.
Select another ECG
lead for processing
(from ECG menu).
Check patient cable
and lead wires and
replace if necessary.
Reapply gel or change
electrodes. Check the
MultiMed pod and
replace if necessary.