INFINITY DELTA/DELTA XL/KAPPA - QUICK REFERENCE GUIDE
CO Problems
CO value is different
form expected
IBP S <#>
IBP D <#>
IBP M <#>
Out of Range (high)
Out of Range (Low)
Please Check Zero
Infinity Delta/Delta XL/Kappa - Quick Reference Guide
Intended for quick reference only.
Refer to Instruction For Use Manual.
Possible Cause
CO SETTINGS are
incorrect. CO injection
takes too long.
Pressure value outside
alarm limits.
Pressure signal out of
measurement range.
The IBP zero stored in
the monitor may not
correspond to the
peripheral device.
Action
Check CO SETTINGS
and input correct
information. Inject over
10-20 Seconds.
Check patient and
take appropriate
action. Access the
Alarm Limits menu
nd change the alarm
limits. Check equip-
ment and replace if
necessary.
Check patient and
treat if necessary.
Access the Pressure
Labels menu and
assign the correct
labels. Check equip-
ment and replace if
necessary.
Zero the transducer.
© 2010 Draeger Medical, Inc.
| 23