INFINITY DELTA/DELTA XL/KAPPA - QUICK REFERENCE GUIDE
NIBP Problems
NBP No Pulsation
MAP appears to be
incorrect
NBP </> alarm limit
NBP measurement
timeout
Infinity Delta/Delta XL/Kappa - Quick Reference Guide
Intended for quick reference only.
Refer to Instruction For Use Manual.
Possible Cause
Weak signal. Monitor
is unable to detect a
sufficient number of
pulsations of adequate
amplitude within two
minutes.
MAP calculated on
displayed SBP and
DBP.
Incorrect limits or
incorrect patient
category.
Measurement lasted
longer than 2 mins
(adult or peds) or
90 sec (neo).
Action
Check the patient and
treat if necessary.
Check the hose and
cuff. Check for proper
size, placement of
cuff.
No action required.
The monitor uses the
step deflation method.
The MAP is determined
first. The DBP and
SBP are calculated
from the MAP.
Check alarm limits
Check patient
category.
Move cuff to limb with
less movement. Ensure
proper cuff size. Adjust
inflation limit. Retry
measurement.
© 2010 Draeger Medical, Inc.
| 21