Portable and Mobile RF Communications Equipment and the KaVo NOMAD PRO 2 device
The KaVo NOMAD PRO 2 Intraoral X-ray System is intended for use in the electromagnetic environment
in which radiated disturbances are controlled. The customer or user of the KaVo NOMAD PRO 2 Intraoral
X-ray System can help prevent electromagnetic interference by maintaining a minimum distance between
portable and mobile RF Communications Equipment and the KaVo NOMAD PRO 2 Intraoral X-ray Systems
recommended below, according to the maximum output power of the communications equipment.
Immunity Test
Rated Maximum
Output Power
(Watts)
0.01
0.1
1
10
100
For transmitters rated at a maximum output power not listed above, the recommended separation distance
d in metres (m) can be determined using the equation applicable to the frequency of the transmitter,
where P is the maximum output power rating of the transmitter in watts (W) according to the transmitter
manufacturer.
NOTE 1 At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies.
NOTE 2 The ISM (industrial, scientific and medical) bands between 150 kHz and 80 MHz are 6.765 MHz to
6.795 MHz; 13.553 MHz to 13.567 MHz; 26.957 MHz to 27.283 MHz; and 40.66 MHz to 40.70 MHz.
50
Recommended Separation Distances Between
IEC 60601
Test Level
Separation Distance According to Frequency of Transmitter
Separation (m)
Separation (m)
80 to 800MHz
150kHz to 80MHz
D=(3.5/V1)(Sqrt P)
D=(3.5/E1)(Sqrt P)
0.117
0.117
0.369
0.369
1.167
1.167
3.689
3.689
11.667
11.667
Electromagnetic Environment
Compliance Level
– Guidance
m
Separation (m)
800MHz to 2.5GHz
D=(7/E1)(Sqrt P)
0.233
0.738
2.333
7.379
23.333