KaVo NOMAD PRO 2 Operator's Manual page 41

Handheld x-ray system for intraoral radiographic imaging
Table of Contents

Advertisement

X-ray Controls and Generator
Leakage technique factors
Maximum air kerma at handgrips and
control panel
Measurement Base of Technique Factors
The kV is measured during pre-pot testing using a calibrated high voltage divider with a
guaranteed accuracy of ±2%. Final performance measurements are made using a NERO
mAx model 8000 X-ray meter from Victoreen. Tube current (mA) is sensed across a series
connected resistor with an accuracy of ±1% and measured using a digital multimeter,
prior to encapsulation; the KaVo NOMAD PRO 2 has no provision for external
measurement of beam current after final manufacture. Exposure time is measured
during the entire exposure, references to 75% rise/fall, using the NERO mAx model 8000
X-ray meter.
Collimator Cone
Minimum source to skin distance
Nominal dose output at cone tip (20cm)
X-ray field size and configuration
Dose Area Product (DAP)
Exposure Time
kV
mA
(milliseconds)
60
2.5
Overall deviation of the air kerma from the values shown does not exceed 40%. The exit
field size at the end of the collimator cone (20 cm from the focal spot) is 6 cm. Therefore,
DAP = mGy × ���� × �
Compliance with Applicable Standards
Standard
Complies with X-ray equipment for dental intra-oral
IEC 60601-2-65
radiography IEC 60601-2-65:2012, KaVo NOMAD
PRO 2.
60kV, 2.5mA, 1.00 sec.
<0.08mGy in 1 hour
21cm (from focal spot to cone tip)
3.41mGy/sec.
6cm diameter circle
Air Kerma
20
40
400
600
1,000
6 cm
2
2
Dose Area Product
(mGy)
0.07
0.14
1.36
2.05
3.41
Title
(mGy*cm
)
2
2
4
39
58
96
Edition
2012
41