KaVo NOMAD PRO 2 Operator's Manual page 31

Handheld x-ray system for intraoral radiographic imaging
Table of Contents

Advertisement

A
& A
: V
lerts
lArms
V
isuAl
Low Battery
Alarm
Incomplete
Exposure
Alarm
System Failure
Alarm
Handset
Failure Alarm
The KaVo NOMAD PRO 2 should not be operated if it has been dropped or if
performance degrades; it should be returned to KaVo for an evaluation.
A
isuAl And
udible
A
udible
5 audible tones
A warning tone
Descending tone
5 audible tones
i
F
ndicAtors
unction
The Recharge Handset alarm
terminates after five seconds
and goes into auto shutdown.
Replace the current handset with
a freshly charged handset.
If the battery voltage is lower than
required for the X-ray exposure, the
device will not allow the exposure.
This alarm activates if the trigger
is released before the timed X-ray
exposure finishes. This condition is
cleared by pressing any button on
the user interface panel, by turning
the power off then on. To avoid this
error, depress the trigger for the
duration of the exposure time.
See
Section 6.2 Troubleshooting
(6.2.8) or an KaVo authorized
service center.
See
Section 6.2 Troubleshooting
(6.2.12) or an KaVo authorized
service center.
/r
esolution
NOTE:
31