Device Symptom
6.2.7 KaVo NOMAD
PRO 2 automatically
shuts down.
6.2.8 System Failure
alarm displayed
on-screen.
6.2.9 There was no
power to KaVo NOMAD
PRO 2 control panel.
6.2.10 When placing
handset on the
charging cradle,
indicator lights
illuminate solid or
flashing red.
6.2.11 The handset did
not seem to be working
when the trigger was
pulled.
Potential Problem
KaVo NOMAD PRO 2
times out after about
three minutes of inactivity.
A different problem exists
if shut-down occurred
during regular activity.
Self-diagnostics, which
automatically run during
X-ray firing, detects a
potential issue.
If pressing Power several
times does not cause the
display to illuminate, the
attached handset is the
likely problem.
A different problem exists
if all handsets appear
fully charged.
If the indicator lights are
solid red, there is an issue
with either the handset or
charger.
If the indicator light
blinks red, the handset is
outside of the operating
temperature range.
The trigger is not
communicating with the
device.
Corrective Action
Manually turn on KaVo NOMAD PRO 2
when you are ready to use the device.
If this condition persists, KaVo NOMAD PRO
2 will require authorized service. See
7.0 Maintenance and
This warning alarm can be cleared
by powering off, then on. If the alarm
redisplays, the KaVo NOMAD PRO 2 will
require authorized service; see
Maintenance and
Repair.
Ensure handset is securely attached.
Replace with a newly charged handset in
order to continue. Use the charging cradle to
determine if the first handset battery can be
recharged or is impaired and must be taken
out of service.
KaVo NOMAD PRO 2 will require authorized
service; see
Section 7.0 Maintenance and
Repair.
Ensure that the compatible components are
being used by checking the labels of both the
handset and charger and verifying that the
correct part numbers are being used.
If correct components are being used, the
handset and charger will require authorized
service; see
Section 7.0 Maintenance and
Repair.
Remove the handset from the charger.
Allow the handset to equilibrate to room
temperature. Place the handset on the
charger. If the error persists, the handset will
require authorized service: see
Maintenance and
Repair.
KaVo NOMAD PRO 2 may require authorized
service; see
Section 7.0 Maintenance and
Repair.