Use right arrow or ENTER and the Inputs & sensors screen occurs. Now set all parameters for CH1
and CH2 (if it is used). The description of all parameters – see [1], chapter Analyser Menu / Inputs
and Sensors Item.
If you use two sensors, then both sensors have to have the same unit.

Measurement Parameter

Use right arrow or ENTER and on the next screen define next balancing parameters. These
parameters are equal with parameters from Ampl. & phase measur. setup - see [1], chapter
Analyser Menu / Ampl. & phase measur. Item.
Additional parameter there is MASS-UNIT parameter, which defines unit of trial mass and of result
balancing masses. When you use the relative trial mass (it means equal 1) then this mass unit is only
formal.
The mass values are always positive. When you want removed the mass then you have to do it on the
opposite side of rotor.

Stop mode Parameter

For the 03.14 FW version and later.
Use right arrow or ENTER and from the next menu select a desired mode of a balancing
measurement stopping.
Adash Ltd., Czech Republic, Tel.: +420 59 6232670, Fax: +420 59 6232671, e-mail: info@adash.cz
For further technical and contact information consult www.adash.net, www.adash.cz
- single measurement: Only a single measurement of the
amplitude, phase and speed is performed. Its result is
stored as the balancing measurement result.
- user stop: The amplitude, phase and speed on-line
measurement cycle is run. Watch the measured phase and
speed values within the measurement cycle. If a correct
balancing speed value is reached and the phase value is
stable, then break the measurement cycle by pressing of
ESC key on the instrument keypad. The last measurement
result is stored as the balancing measurement result.
15
A4300 - VA3 - Balancing