Proper Placement Of O-Ring - Beckman Coulter GH-3.8 Instructions For Use Manual

Swinging bucket rotor. in beckman coulter allegra 6 series, gs-6 series, spinchron series, and gp series centrifuges
Table of Contents

Advertisement

GH-3.8 Rotor
Loading the Rotor
Install the covers as follows:
1
Make sure that the bucket and the bucket cover surface are clean, dry, and undamaged.
2
Make sure that the O-ring (961648) is in good condition and lightly coated with silicone vacuum
grease (335148)
3
Place the O-ring on the ledge inside the bucket (see
Figure 10 Proper Placement of O-Ring
4
Place the cover on the bucket with the latches perpendicular to the pin sockets (see
5
Snap the latches down and secure them under the bucket rim.
Figure 11 Securing Bucket Cover Latches
Snap Down
Pin Socket
1-24
Insert
O-ring here
Figure
10).
Cover
Spring
Latch
Bucket
Figure
11).
PN GS6-TB-003MD