Figure 2-2 Dut To Power Meter Connection - Keysight Technologies 8481A Operating And Service Manual

Coaxial power sensors
Table of Contents

Advertisement

Recommended 50 MHz, 50 Ohm Type-N (f) to 75
Ohm Type-N (m):
Recommended 50 MHz, 30 dB Attenuator Pad
APC-3.5mm to Type-N:
Procedure
Figure 2-1
1 Connect the DUT (8483A/1D/5D/7D) to the power meter as shown in the
Figure 2-2
2 Warm up the DUT for approximately 30 minutes.
Keysight 8480 Series Operating and Service Manual
Power
meter
DUT
Zero set performance verification equipment setup
following figure.
Power meter
Channel A
DUT to power meter connection
PC
Power sensor
Power ref
Power sensor cable
General Information
11852B adapter
(used with 8483A)
11708A adapter
(used with 8481/5/7D)
2
45