Power Meter Calibrations

Follow the calibration procedures given in your power meter manual. The
E4413A power sensor is fitted with 3.5 mm (m) connectors as standard.
To convert the 3.5 mm (m) connector for calibration, an adapter (3.5 mm
(f) to Type- N (m)) is included with the power sensor. See
Figure 3
The 3.5 mm to Type-N adapter is intended for the use of 1 mW, 50 MHz power reference of
N O T E
the power meter only. Its function as a calibration reference may be compromised if it is
used for other purpose.

Operating Instructions

The E4412A and E4413A power sensors are compatible ONLY with the
newer E44XX- Series power meters. They are NOT compatible with the
earlier 430- Series, E1416A, or 70100A power meters. To operate the power
sensor, refer to the operating instructions in the Keysight E44XX- Series
Power Meter User's Guide.
14
E4413A power sensor with adapter
E4412A and E4413A Operating and Service Guide