Configure Subsystem; Configure[1]|2|3|4[:Scalar][:Power][:Ac] - Keysight Technologies U2040 X-Series Programming Manual

Wide dynamic range power sensors
Table of Contents

Advertisement

6

CONFigure Subsystem

CONFigure[1]|2|3|4[:SCALar][:POWer][:AC]

Syntax
CONFigure[1]|2|3|4[:SCALar][:POWer][:AC]
[<expected_value>[,<resolution>[,<source_list>]]]
CONFigure[1]|2|3|4[:SCALar][:POWer][:AC]?
Description
This command is used on the specified CALCulate block to set:
– the expected power level of the measurement.
– the resolution of the measurement.
– the channel on which the measurement will be made.
The values shown in this command are those that are tested.
The query returns the present configuration of the specified CALCulate block.
Parameters
Item
expected_value
resolution
source_list
[a] The mnemonic DEF means DEFault. This is not equivalent to the DEFault parameter used in the command
subsystems. The parameters must be entered in the specified order. If parameters are omitted, they default from the
right. The parameter DEFault is used as a placeholder. Specifying DEF leaves the parameter value unchanged.
[b] When SENS:AVER:COUN:AUTO is enabled, the resolution setting affects the number of averages applied. Refer to
"Auto-averaging mode" on page
Example
CONF DEF,2,(@1)
118
Range of values
Description/Default value
A numeric value for the expected power level. The units of
sensor-dependent
measurement are dBm and W. The default units are defined by
[a]
DEF
UNIT:POWer.
[b]
1 to 4
A numeric value for the resolution. If unspecified, the current
1.0, 0.1, 0.01, 0.001
resolution setting is used.
[a]
DEF
The channel which the command is implemented on.
If unspecified, the current setup is used. If the ratio or
(@1)
difference measurement was set up, it will default to
Channel A.
44
for more information.
Configures CALCulate1 to measure the power of Channel A,
using the current range and a resolution setting of 2.
Keysight U2040 X-Series Programming Guide