8480 Series Power Sensor Specifications; Table 1-3 Typical Measurement Uncertainties Of Calibration Factor (Cf) Data Printed On The Power Sensor - Keysight Technologies 8481A Operating And Service Manual

Coaxial power sensors
Table of Contents

Advertisement

8480 Series Power Sensor Specifications

These specifications are valid with EPM and EPM-P Series of power meters.
NOTE
The 8480 series thermocouple and diode power sensors provide accuracy,
stability, and SWR over a wide range of frequencies (100 kHz to 50 GHz) and
power levels (-70 dBm to +44 dBm)
Table 1-3
Frequency
100 kHz to 10 MHz
10 MHz to 30 MHz
30 MHz to 500 MHz
500 MHz to 1.2 GHz
1.2 GHz to 6 GHz
6 GHz to 14 GHz
14 GHz to 18 GHz
18 GHz to 26.5 GHz
26.5 GHz to 33 GHz
33 GHz to 34 GHz
34 GHz to 35 GHz
35 GHz to 40 GHz
40 GHz to 45 GHz
45 GHz to 50 GHz
[a] These uncertainties only apply to Option 033.
Keysight 8480 Series Operating and Service Manual
Typical measurement uncertainties of Calibration Factor (CF)
data printed on the power sensor
8481A
-
0.81
0.78
0.78
0.92
1.16
1.59
-
-
-
-
-
-
-
25°C ± 3°C
8482A
8483A
0.87
1.59
0.8
1.39
0.8
1.41
0.8
1.41
0.9
1.46
-
-
-
-
-
-
-
-
-
-
-
-
-
-
-
-
-
-
Introduction
1
8485A
-
-
1.29
1.26
1.35
1.61
1.77
2.47
[a]
3.35
-
-
-
-
-
23

Hide quick links:

Advertisement

Table of Contents