Figure 2-3 Power Linearity Performance Verification Equipment Setup - Keysight Technologies 8481A Operating And Service Manual

Coaxial power sensors
Table of Contents

Advertisement

– power splitter (11667A-Option 001)
– amplifier
– step attenuators (8494H and 8496H)
– attenuator/switch driver (11713B)
– 50 MHz, 30 dBm attenuator pad (11708A)
Procedure
1 Turn on the signal generator and power meter. Connect the DUT (848xA) to
2 Zero and calibrate the DUT using the reference sensor with the power meter.
3 Connect the power splitter (11667A option 001) to the RF output of the signal
Figure 2-3
4 Set the continuous wave signal frequency of the signal generator, DUT, and
Keysight 8480 Series Operating and Service Manual
the channel A and the reference sensor (N8481A) to the channel B of the
power meter (E4416/7A). Allow them to warm up for approximately an hour.
For 8481/5/7D models, during the zero and calibration please use the 50 MHz,
30 dBm attenuator pad (11708A).
generator (N5182A). The equipment setup is as shown in the following figure.
Signal generator
Power splitter
DUT
Power meter
Power linearity performance verification equipment setup
reference sensor to 50 MHz. Set DUT to AVERAGE ONLY mode.
10 dB step attenuator
1 dB step attenuator
Reference
sensor
General Information
Amplifier
Attenuator/
switch driver
(Optional)
2
47