7 The startup/shutdown of a pump

7.4.1 Conservation

If the micro annular gear pump operates at irregular intervals or for other
reasons should be put out of operation for a longer period, it should, after
service and flushing procedure (see chapter 7.3), be filled in with a suitable
conservation liquid.
The conservation liquid may be selected from the table 35 depending on the
duration of the standstill and the resistance of the pump to the manipulated
liquid (table 34). The indicated conservation liquids are simple
recommendations and should therefore be checked by the user as to their
compatibility and suitability. The figure 33 presents the diagram of conservation
liquid selection.
Remark: This diagram is repeated as a part of the figure 32 »Shutdown
procedure«.
After the cleansing procedure the pump should be filled with a suitable
conservation agent. You will find a choice of possible conservation liquids in
the table 35.
figure 33
Selection of the conservation liquid (CL)
Last update: January 2016
Selection of conservation
liquid (CL)
CL
no
adequate
yes
Is the CL
no
compatible
with pump
yes
Is the CL
compatible
no
with the
last FL ?
yes
Technical data subject to change without prior notice!
Operating manual mzr-2542 / mzr
see table 35 (Selection of conservation liquid)
see table 34 (resistance of the sealing materials)
-
2942
51