7 The startup/shutdown of a pump
7.4

Shutdown of the micro annular gear pump

During the shutdown of the pump the following steps should be followed
 Flush the pump with a filtered and particle-free flushing liquid (solvent)
 After the flushing procedure decrease speed of the pump to 0 rpm
 Fill the pump with a suitable conservation liquid (see chapter 7.4.1)
 Remove the pump from the system (see chapter 7.4.2)
By proceeding as shown in the diagram (see figure 32) you may prepare the
pump for a longer standstill.
Last update: January 2016
(see chapter 7.3)
Technical data subject to change without prior notice!
Operating manual mzr-2542 / mzr
-
2942