Power connection to screw terminals (standalone unit)
1.
Loosen the lid clip and unscrew the lid
of the enclosure.
2.
Unscrew and lift up the digital display.
3.
Connect the wires to the terminals:
polarity is not important. (Terminal is
removable.)
4.
To use the Alarm Output, connect the
wires of an optional input to the Alarm
Output terminals: polarity is not
important. (Terminal is removable.)
5.
Ground the instrument according to
1
local regulations
6.
Tighten the gland to form a good seal.
alarm output
solid-state switch
Operation
Digital display
Selected in Mode 13: displays either the measured value, logical level, mode number and
numerical value for the selected mode, or electronics temperature.
Measured value display
• default display at startup
2
• value (in counts
1.
The usual PROFIBUS PA recommendation is to ground the shield on both the device side and the cable
side. In some cases it may be preferable to ground one side only, to avoid ground loops.
2.
A dimensionless value generated from the inverse of frequency.
3.
Switch function (open/closed) depends on setting at Mode 24.
Page EN-6
Pointek CLS200/300 (Digital) – QUICK START MANUAL
.
test
input
sensor
PROFIBUS PA
PA+
PA–
power cable
) when Sensor (2) selected
removable
terminal block
for alarm output/
solid-state
switch
7.
Replace the digital display.
reed
8. To adjust the transmitter locally,
contact
use the keypad. (See
button keypad
Quick Reference table on page 8).
After adjustment, replace the
enclosure lid and secure the lid clip.
3
down arrow
(switch open
)
3
bar graph
digital display
removable
terminal block:
for PROFIBUS
PA/power
connection
lid clip
gland
power
cable
Setup using 3-
on page 7 and the
sensor uncovered
7ML19985XA82