3.
Instrument components and specifications

3.3 Technical data

General
Approvals:
Operating temperature range:
Temperature range during storage:
Relative humidity:
Humidity during storage:
Section thickness range:
Section thickness settings:
Specimen feed:
Vertical stroke:
Max. sectioning range without retraction: 69 mm (without specimen orientation at 1 µm)
Max. sectioning range with retraction:
The specimen retraction can be turned
off manually:
Dimensions and weight
Width (including handwheel):
Depth (including waste tray):
Height (total):
Working height (knife blade):
Weight (without accessories):
Optional equipment and optional accessories
Object orientation (option)
horizontal:
vertical:
Rotating range:
Trimming increments:
Repositioning of knife holder base
north-south:
14
The instrument-specific marks are located on the
rear panel of the instrument next to the name plate.
+10 °C to +35 °C
+5 °C to +55 °C
max. 80 % non-condensing
< 80 %
1.0 - 60.0 µm
from 1.0 - 10.0 µm in 1.0 µm increments
from 10.0 - 20.0 µm in 2.0 µm increments
from 20.0 - 60.0 µm in 5.0 µm increments
approx. 24 mm, ± 2 mm
70 mm
62 mm
200 µm
413 mm
618 mm
305 mm (with tray on the hood)
168 mm (measured from the table)
approx. 37 kg
8 °
± 90°
10 µm, 50 µm
± 25 mm
Instruction manual V 1.4 – 02/2010