Installation; Requirements For Installation; Requirements Of The Installation Location; Transport Of The Unit - gbo Medizintechnik ULTRATHERM 1008 Instructions For Use Manual

Short-wave therapy unit
Table of Contents

Advertisement

ULTRATHERM 1008
3
3.1
3.2
3.3
© gbo Medizintechnik AG

Installation

Requirements for installation

Before the unit can be installed and put into operation, certain requirements
must be fulfilled in the building where the unit is to be operated!
If the unit cannot be installed immediately after delivery, the unit and its
external components or accessory elements must be stored in their orig-
inal packaging in a dry place!
Do not store or operate the unit in a dusty environment!

Requirements of the installation location

The unit must be installed so that there is no danger to the patient, the
operator or other persons!
Therefore, you must read the safety instruction in Chapter 2 and the fol-
lowing information!
By selecting a suitable location for setting up the unit or by means of
structural measures, contact during the treatment by the personnel or
the patient with conductive materials that are earthed or have a high ca-
pacity to earth must be prevented (e.g. heating pipes, water faucets,
metal chairs, metal beds or other earthed devices).
The unit must be set up so that the (normal) release of electromagnetic
radiation during operation does not hinder the function of other devices
or data media. The minimum distance to other devices or their power
supplies or data transfer lines is 6 meters! Please note that the radiation
can easily pass walls, ceilings and floors.
The room and the installation location must be large enough so that the
unit can be operated from the front even if the electrodes are positioned
inconveniently.

Transport of the unit

Measures concerning the transport of the unit from the manufacturer to the
operator are based on the individual circumstances and are defined in the
general terms of business.
In the event of subsequent transport of the unit, the dealer or the operator is
responsible for the unit and for compliance with the safety and accident pre-
vention regulations.
13
Version 1.7

Hide quick links: