ULTRATHERM 1008
2.3
© gbo Medizintechnik AG

Personal safety

In case of improper or unauthorized use of the unit, the operator, the pa-
tient or other persons may be subjected to the danger of electric shock
due to high voltage produced by the unit, the danger of influence on ac-
tive implantations by magnetic fields produced by the unit and the dan-
ger of being burned due to erroneously positioned electrodes or false
parameters such as the duration of treatment, power output or operating
mode!
Before operating the unit, please read this instruction manual carefully
and observe the information contained therein!
Pay special attention to the list of contraindications at the end of the in-
struction manual!
( Annex B Contraindications)
Before operating the unit each time, check whether
the unit has been correctly connected to the building's power supply,
the unit has been set up so that it is free-standing and the patient is not
in direct contact with metal objects such as heating radiators, metal beds
or other equipment,
the insulation of the power supply and electrode connection cables is not
damaged,
the electrode connection cable is connected properly and is not cross-
routed (which may cause capacitive short circuits),
only accessories (cables, electrodes) approved by the manufacturer are
connected,
the patient to be treated (and the personnel) have removed all electric
devices (e. g. hearing aids, electrotherapy electrodes, mobile tele-
phones) and all conductive objects (e. g. rings, chains, watches, earrings
or other jewelry, eyeglasses) and that they are not in the immediate vi-
cinity of the unit,
the patient is in a composed state and the bodily areas to be treated are
dry on the exterior,
the electrodes are positioned according to the doctor's instructions (to be
checked by the doctor or physiotherapist if applied by assisting person-
nel),
there is no danger of unwanted local warming due to electrode con-
strictions and
no other persons are located within 2 meters of the unit!
9
Version 1.7