ULTRATHERM 1008
9
9.1
9.2
© gbo Medizintechnik AG
Technical specifications
Transport and storage conditions
Ambient temperature **
Relative humidity
Air pressure
Operating data and ratings
Dimensions *
Weight *
Ambient temperature **
Relative humidity
Air pressure
Supply voltage
Frequency
Power consumption
Line fuses
Load impedance
HF nominal power
Power setting
Power indication
Working frequency
Radiation type
Pulse duration
Pulse frequency
Treatment duration
Applied part
Type of applied part
Protection class
Equipment class
*
without electrodes, electrode arms and electrode cables
**
If the unit is operated after cooling to below the freezing point, mal-
functions or defects may occur due to formation of condensation
water !
+ 5 °C to + 50 °C
< 90 %
500 hPa to 1060 hPa
420 x 970 x 410 mm (W x H x D)
56 kg
+ 10 °C to + 40 °C
30 % to 75 %
700 hPa to 1060 hPa
115 / 230 V
50-60 Hz
1400 VA
2× T 16A L 250V, glass fuse 6.3×32mm
50
1000 W ±20% pulsed mode (peak)
400 W ±20% continuous mode
in 10-Watt increments (CW mode),
in 50-Watt increments (pulsed mode)
Peak power and effective power
27.12 MHz ± 0.6 %
non-ionizing
200 – 600 s, ± 2 %
10 – 300 Hz, ± 2 %
1 – 30 min, ± 0.5 %
Applicators
BF
I
IIb acc. to MDD
41