Error messages

Both operator errors and internal errors in the unit are displayed as an error
code with text information.
Operator errors
E01: Please check electrodes and cables
Cause: An optimum tuning point could not be found; therefore, the appli-
cators, their ESD and the correct connection to the electrode cable
should be checked again.
E02: Matching...Please wait!
Cause: For reasons of safety the effective output must not be set higher
than 100 W, as long as the tuning process is still in progress. The rea-
son for this is that if tuning takes place at a higher effective output, heat
can be clearly felt.
Internal errors
With the exception of temperature errors, the treatment must be discontinued
immediately in case of internal errors in the unit. The unit can only be switched
off and on again by means of the power switch. If the error message appears
again, make a note of the error code, switch the unit off again and notify your
authorized service partner.
In case of display failure or other obvious defects, switch the unit off
immediately by means of the power switch and notify your authorized
service partner!

Repairs

Do no perform unauthorized repairs under any circumstances!
All service work (safety inspections, repairs) may be performed only by service
personnel authorized by the manufacturer!

Disposal of equipment and accessories

According to the WEEE Directive 2012/19/EU (waste electrical and electronic
equipment) this device must not be disposed of with the domestic waste. The
device must be returned to the manufacturer for disposal. The manufacturer is
committed to guarantee the disposal of all devices marketed. This is also indi-
cated by the WEEE sign (crossed out waste container).