Do not operate the unit for an extended period with no load (without
a patient), especially in coil (induction field) mode! When operating the
unit without power output, induction field electrodes could be
destroyed due to overheating!
Pay attention to the routing of the electrode connection cables. These
must always be in the air and must never lie on metal surfaces!
Do not treat Patients on metal chairs or metal beds.
Keep chip cards, magnetic cards, audio and video cassettes and other
data media that are susceptible to interference away from the unit!
Clean and disinfect the unit only when the power supply is deactivated
(power switch off, power plug disconnected)!
Clean and disinfect the unit only by means of disinfection by wiping!
Disinfecting by spraying can damage the unit due to penetrating mois-
ture!
Never clean the unit with abrasives, disinfectants or solvents that could
scratch the housing or damage the unit!
Never perform unauthorized service work!
All service work (safety inspections and repairs) must be performed only by
service technicians who have been authorized by the manufacturer!
During service and maintenance work it is not allowed to treat a patient!
The device must not be used as a step or a seat!
ULTRATHERM 1008
Version 1.7