Switching on the unit
Switch on the power switch.
After switching on the unit by means of the power switch, a short self-test is
conducted.
In case of an internal error in the unit, an error code is displayed in the text
output field, to assist the technicians in locating the error. In this case, please
contact your authorized dealer.
Setting the parameters
Parameters can be set in three ways:
1. Direct input of parameters
2. Selection of a disorder in the indications menu
3. Selection of a saved program in the program menu.
Proceed as follows, depending on the desired method for setting the parame-
ters:
Direct input of parameters
According to the information in the indications table in Annex A, set the
parameters for duration of treatment, operating mode and frequency
(only for pulsed mode) as follows:
1. Select desired parameter by pressing the corresponding function
key (min., Mode, Hz) in the main menu.
2. Set the selected parameter by turning the multi-function rotary switch
to the appropriate position.
3.
Accept the set value by pressing the multi-function rotary switch