How To Open The Door - Scilogex DM0636 User Manual

Multi-purpose centrifuge
Table of Contents

Advertisement

USER MANUAL
E-02
Door fault
E-03
ROTOR ID
E-04
Temp sensor fault
E-06
Set wrong parameter
E-08
Chamber over hot
E-09
imbalance
E-10~86

10.2 How to open the door

10.2.1 In the case of power on
CAUTION
The door just can be opened while the power on and rotor stops rotating.
(1)Turn on the POWER switch, the door lock will release automatically.
(2)The door lock will release automatically once the operation finished.
(3)It is available to release the door by pressing key
10.2.2 In the case of power outage
The door cannot be opened automatically if there is a power outage. It is available to be opened manually.
(1)Ensure if the rotor has stopped rotating.
Listen carefully to ensure no rotating sound can be heard.
(2)Insert a screw driver into the hole to open the door.
· The door opened in running.
· Press the key
door opening.
· The centrifuge can not identify the
rotor ID.
· the connection fault.
· the sensor fault.
· The setting parameter exceed the
allowable range.
· The air inlets are blocked.
· cooling fan is damaged.
·
the allowable imbalance is
exceeded.
·some wrong in the drive system
· Read the service manual.
Table 10-1 Possible problems and solutions
· Close the door immediately.
· Close the door , and then start to
while the
operate.
·Reconfirm the ID code of the rotor and
make a correct selection.
· reconnect
· Replace temperature sensor
· Modify the parameter value.
· Clean air inlets.
· Replace the cooling fan.
·Balance the sample
·checked by professional person.
· Contact with service center.
once the rotor stops.