Routine Inspection; Troubleshooting; Possible Problems And Solutions - Scilogex DM0636 User Manual

Multi-purpose centrifuge
Table of Contents

Advertisement

USER MANUAL
rings may fall off or aging after several autoclaving, need to be replaced or re-installed.
Seal ring 1 2
9.3.2 Replacement methods
(1)Clean the seal ring slot with neutral detergent solution and make it dry.
(2) Evenly coated with glue (501) in the seal ring slot and keep the seal ring into slot, press evenly to make
it contact enough with the slot bottom and bond firmly.
(3)Place for 20 minutes and waiting for the glue to completely solidified.

9.4 Routine inspection

(1) Check that if the centrifuge is on a firm、flat and level table, ensure the four feet stand on the table
firmly.
(2) Check if the centrifuge grounded properly: Use multi-meter to check if is short circuit between the
power cord grounding pin and the motor shaft. If yes, indicating grounded properly; if is open circuit, need
to check failure reason first and make troubleshooting before use.

10. Troubleshooting

10.1 Possible problems and solutions

This centrifuge has a self-diagnostic function. If a problem occurs, an error/warning code will be displayed
on the time display screen and the operator can determine the malfunction with the warning code below.
Symptom
Nothing
appears
on
screen when the POWER is
turned on.
3
Figure 9-1
Seal rings of rotor
Causes
the
· Building power circuit breaker
trips.
· the fuse was blown out.
Seal ring 2
Solutions
· Remove the trouble and turn on the
POWER.
· Replace the fuse.
14
3