stuck between the rotor and the shaft. Double check and clean it if necessary.
Rotate the rotor slightly with your fingers to check if the rotor vibrates. If so reinstall the rotor again.
Rotate the nut clockwise using the wrench to tighten the rotor to the shaft firmly.
Close the rotor lid, firmly tighten clockwise the lid to the rotor and ensure is in position. Close the door
and then start running.
The method of removing the rotor is as same as the above mentioned by turning the locking nut
counterclockwise.
8.1.2
Set the operation parameters
Press the
key to select required parameters. The parameter can be modified when the parameter is
flashing. Clockwise rotate the parameter key
parameter key
to decrease parameter value. Parameter key
faster. The minimum speed increment is 100 rpm, the minimum time increment is 1 second.
(1)Set the speed
Press the key
When the speed is selected, the speed symbol will flash the speed value.
The minimum speed value you can set 300rpm, the minimum increment is 100rpm.
Rotate parameter key
anti-clockwise to decrease speed value.
You can speed-up set the speed value by rotating parameter key
There is a circulating function to increase/decrease the speed values. Rotate the parameter key
to change settings from small → large → maximum → minimum. Rotate the
clockwise
parameter key anti-clockwise
(2)Set the time
Press key
, time value flashes in the time setting mode.
Rotate the parameter key
When time displays HD, this is a continuous running mode.
(3) Set acceleration and deceleration
Press key
, acceleration value flash, press the
change from 1 to 9, then from 9 to1.
1 acceleration: the slowest acceleration; 9 acceleration: the fastest acceleration.
Press key
, deceleration value flash, press the
change from 0 to 9, then from 9 to 0.
0 deceleration: free braking;
1deceleration: the slowest deceleration; 9deceleration: the fastest deceleration.
until the speed rpm is displayed.
clockwise to increase speed value, rotate the parameter key
to change settings from large → small → minimum → maximum.
to set running time from 10 seconds to 99 minutes.
to increase parameter value; counter-clockwise rotate the
rotate faster, parameter value increase
again, the value will increasing.the value will
again, the value will increasing.the value will